Scope of Acreditation of Defence Institute „Professor Tsvetan Lazarov" Central Laboratory for measuring equipment
|
REPUBLIC OF BULGARIA Executive Agency Bulgarian Accreditation Service Signatory to the Multilateral Agreement on the mutual recognition of the EA in this field |
Translation from Bulgarian |
ORDER
No. А 594
Sofia, 18.12.2019
Pursuant to Art. 10, para. 1, item 3 and art. 30, para 1 of the Law on the national accreditation of conformity assessment bodies and item 4.3.7 of the Accreditation Procedure BAS QR 2 in connection with an opened procedure with Ref. No. 8/7 LC/RO/13.08.2019, Declaration for refining the scope of accreditation applied for extension of scope Ref. No 8/7 LC/RO/6/B/28.10.2019, Report of on-site assessment Ref. No. 8/7 LC/RO/5/B/28.10.2019, Declaration for Refinement of the scope of accreditation applied for extension Ref. No. 8/7 LC/RO/7/B/06.11.2019, Standpoint of the Accreditation Commission 8/7 LC/RO/8/B/05.12.2019
I HEREBY EXTEND THE SCOPE OF ACREDITATION
of
DEFENCE INSTITUTE „Professor Tsvetan Lazarov"
CENTRAL LABORATORY FOR MEASURING EQUIPMENT
Management address and address of the laboratory:
1592 Sofia, 2 Prof. Tsvetan Lazarov Blvd.
To perform calibration of:
№ |
Type of measuring instrument |
Measured value, unit of measure
|
Measuring range |
Uncertainty of measurement
|
Calibration method | ||
1. |
DC ammeters |
Direct |
from 0 µА to 10 µА |
from 0,023 µА to 0,029 µА |
РПК 702-12 Direct method |
||
from 10 µА to 100 µА |
from 0,029 µА to 0,081 µА |
||||||
from 100 µА to 1 mА |
from 0,081 µА to 0,000 3 mА |
||||||
from 1 mА to 100 mА |
from 0,000 3 mА to 0,019 mА |
||||||
from 100 mА to 1 А |
from 0,019 mА to 0,000 3 A |
||||||
from 1 A to 10 А |
from 0,000 3 A to 0,005 A |
||||||
from10 A to 20 А |
from 0,005 A to 0,007 A |
||||||
2.
|
AC ammeters
|
Alternating
|
At a frequency |
|
|||
from 10 µA to 100 µA |
from 0,058 µA to 0,20 µA |
РПК 702-12 Direct method |
|||||
от 100 µA до 1 mA |
from 0,20 µA to 0,001 mА |
||||||
from 1 mA to 10 mA |
from 0,001 mA to 0,007 mA |
||||||
from 10 mA to 100 mA |
from 0,007 mA to 0,072 mA |
||||||
from 100 mA to 1 A |
from 0,072 mA to 0,001 A |
||||||
from 1 A to 10 A |
from 0,001 A to 0,019 A |
||||||
from 10 A to 20 A |
from 0,019 A to 0,03 A |
||||||
At a frequency from 400Hz to 1kHz |
|
||||||
from 10 µA to 100 µA |
from 0,058 µA to 0,20 µA |
РПК 702-12 Direct method |
|||||
от 100 µA to 1 mA |
from 0,20 µA to 0,001 mА |
||||||
from 1 mA to 10 mA |
from 0,001 mA до 0,007 mA |
||||||
from 10 mA to 100 mA |
from 0,007 mA to 0,072 mA |
||||||
from 100 mA to 1 A |
from 0,072 mA to 0,001 A |
||||||
from 1 A to 10 A |
from 0,001 A to 0,019 A |
||||||
from 10 A to 20 A |
from 0,019 A to 0,03 A |
||||||
At a frequency from 1kHz to 5kHz |
|
||||||
from 10 µA до 100 µA |
from 0,22 µA до 0,74 µA |
РПК 702-12 Direct method |
|||||
from 100 µA to 1 mA |
from 0,74 µA to 0,007 4 mА |
||||||
from 1 mA to 10 mA |
from 0,007 4 mA to 0,1 mA |
||||||
from 10 mA to 190 mA |
from 0,1 mA to 1,3 mA |
||||||
3. |
DC calibrators |
Direct current Ampere (А) |
from 10 µА to 100 µА |
from 0,035 µА to 0,087 µА |
РПК 702-9 Direct method
|
||
from 100 µА to 1 mА |
from 0,087 µА to 0,000 9 mА |
||||||
from 1 mА to 10 mА |
from 0,000 9 mА to 0,009 mА |
||||||
from 10 mА to 100 mА |
from 0,009 mА to 0,09 mА |
||||||
from 100 mА to 1 А |
from 0,09 mА to 0,001 А |
||||||
from 1 А to 10 А |
to 0,001 А to 0,009 А |
||||||
from 10 А to 20 A |
from 0,01 А to 0,04 А |
||||||
4.
|
AC calibrators
|
Alternating current Ampere (А)
|
At a frequency from 50Hz to 400Hz |
|
|||
from 10 µA to 100 µA |
from 0,22 µA to 0,3 µA |
РПК 702-9 Direct method |
|||||
from 100 µA to 1 mA |
from 0,3 µA to 0,002 mА |
||||||
from 1 mA to 10 mA |
from 0,002 mА to 0,026 mА |
||||||
from 10 mA to 100mA |
from 0,026 mА to 0,2 mА |
||||||
from 100 mA to 1 A |
from 0,2 mА to 0,002 А |
||||||
from 1 A to 10 A |
from 0,002 А to 0,03 А |
||||||
from 10 A to 20 A |
from 0,03 А to 0,04 А |
||||||
At a frequency from 400Hz to 1kHz |
|
||||||
from 10 µA to 100 µA |
from 0,22 µA to 0,3 µA |
||||||
from 100 µA to 1 mA |
from 0,3 µA to 0,002 mА |
||||||
from 1 mA to 10 mA |
from 0,002 mА to 0,026 mА |
||||||
from 10 mA to 100mA |
from 0,026 mА to 0,2 mА |
||||||
At a frequency from 1kHz to 5kHz |
|
||||||
from 10 µA to 100 µA |
from 0,61 µA to 0,65 µA |
||||||
from 100 µA to 1 mA |
from 0,65 µA to 0,003 mA |
||||||
from 1 mA to 10 mA |
from 0,003 mA to 0,047 mA |
||||||
from 10 mA to 100 mA |
from 0,047 mA to 0,4 mA |
||||||
from 100 mA to 190 mA |
from 0,4 mA to 3,7 mA |
||||||
5. |
Voltmeters for DC voltage |
DC voltage Volt (V) |
from 0 mV to 10 mV |
from 0,012 mV to 0,020 mV |
|||
from 10 mV to 100mV |
from 0,020 mV to 0,030 mV |
||||||
from 100 mV to 1 V |
from 0,030 mV to 0,000 1 V |
||||||
from 1 V to 10 V |
from 0,000 1 V to 0,001 V |
||||||
from 10 V to 100 V |
from 0,001 V to 0,005 V |
||||||
from 100 V to 1000 V |
from 0,005 V to 0,12 V |
||||||
6.
|
Voltmeters for AC voltage
|
AC voltage Volt (V)
|
At a frequency from 50Hz to 400Hz |
|
|||
from 10 mV to 100 mV |
from 0,058 mV to 0,21 mV |
РПК 702-12 Direct method |
|||||
from 100 mV to 1 V |
from 0,21 mV to 0,000 4 V |
||||||
from 1 V to 10 V |
from 0,000 4 V to 0,004 V |
||||||
from 10 V to 100 V |
from 0,004 V to 0,057 V |
||||||
from 100 V to 1000 V |
from 0,057 V to 0,59 V |
||||||
At a frequency from 400Hz to 1kHz |
|
||||||
from 10 mV to 100 mV |
from 0,058 mV to 0,21 mV |
||||||
from 100 mV to 1 V |
from 0,21 mV to 0,000 4 V |
||||||
from 1 V to 10 V |
from 0,000 4 V to 0,004 V |
||||||
from 10 V to 100 V |
from 0,004 V to 0,057 V |
||||||
from 100 V to 1000 V |
from 0,057 V to 0,59 V |
||||||
|
|
|
At a frequency from 1kHz to 10kHz |
|
|||
from 10 mV to 100 mV |
from 0,058 mV to 0,21 mV |
||||||
from 100 mV to 1 V |
from 0,21 mV to 0,000 4 V |
||||||
from 1 V to 10 V |
from 0,000 4 V to 0,004 V |
||||||
from 10 V to 100 V |
from 0,004 V to 0,057 V |
||||||
from 100 V to 190 V |
from 0,057 V to 0,084 V |
||||||
At a frequency from 10kHz to 50kHz |
|
||||||
from 10 mV to 100 mV |
from 0,070 mV to 0,31 mV |
||||||
from 100 mV to 1 V |
from 0,31 mV to 0,0008 V |
||||||
from 1 V to 10 V |
from 0,0008 V to 0,013 V |
||||||
from 10 V to 19 V |
from 0,013 V to 0,018 V |
||||||
7. |
DC Voltage Calibrators |
DC voltage Volt (V)
|
from 0 mV to 10 mV |
from 0,004 mV to 0,005 mV |
РПК 702-8 Direct method |
||
from 10 mV to 100mV |
from 0,005 mV to 0,01 mV |
||||||
from 100 mV to 1 V |
from 0,01 mV to 0,000 054 V |
||||||
from 1 V to 10 V |
from 0,000 054V to 0,00046 V |
||||||
from 10 V to 100 V |
from 0,000 46 V to 0,005 9 V |
||||||
from 100 V to 1000 V |
from 0,005 9 V to 0,064 V |
||||||
8.
|
AC Voltage Calibrators
|
AC voltage Volt (V)
|
At a frequency from 50Hz to 400Hz |
|
|||
from 10 mV to 100mV |
from 0,053 mV to 0,12 mV |
РПК 702-8 Direct method |
|||||
from 100 mV tо 1 V |
from 0,12 mV to 0,001 V |
||||||
from 1 V to 10 V |
from 0,001 V to 0,01 V |
||||||
from 10 V to 100 V |
from 0,01 V to 0,10 V |
||||||
from 100 V to 1000 V |
from 0,10 V to 0,97 V |
||||||
At a frequency from 400Hz to 1kHz |
|
||||||
from 10 mV to 100mV |
from 0,053 mV to 0,12 mV |
РПК 702-8 Direct method |
|||||
from 100 mV to 1 V |
from 0,12 mV to 0,001 V |
||||||
from 1 V to 10 V |
from 0,001 V to 0,01 V |
||||||
from 10 V to 100 V |
from 0,01 V to 0,10 V |
||||||
from 100 V tо 1000 V |
from 0,10 V to 0,97 V |
||||||
At a frequency from 1kHz to 10kHz |
|
||||||
from 10 mV to 100mV |
from 0,10 mV to 0,37 mV |
РПК 702-8 Direct method |
|||||
from 100 mV to 1 V |
from 0,37 mV to 0,059 V |
||||||
from 1 V to 10 V |
from 0,059 V to 0,072 V |
||||||
from 10 V to 100 V |
from 0,072 V tо 0,21 V |
||||||
from 100 V tо 190 V |
from 0,21 V to 0,34 V |
||||||
At a frequency from 10kHz to 50kHz |
|
||||||
from 10 mV to 100mV |
from 0,10 mV to 0,37 mV |
РПК 702-8 Direct method |
|||||
from 100 mV to 1 V |
from 0,37 mV to 0,059 V |
||||||
from 1 V to 10 V |
from 0,059 V to 0,072V |
||||||
from 10 V to 19 V |
from 0,072V до 0,86 V |
||||||
9.
|
Digital R, L, C meters
|
Inductance Henry (Н)
|
At a frequency 1 kHz |
|
|||
from 10 µН to 100 µН |
from 0,25 µН to 0,25 µН |
РПК 702-3 Direct method |
|||||
from 100 µН to 1 mН |
from 0,25 µН to 0,001 mН |
||||||
from 1 mН to 10 mН |
from 0,001 mН to 0,007 mН |
||||||
from 10 mН to 100 mH |
from 0,007 mН to 0,07mН |
||||||
from 100 mН to 1 H |
from 0,07 mН to 0,005 8 H |
||||||
Capacitance Farad (F)
|
At a frequency 1 kHz |
|
|||||
from 1 pF to 10 pF |
from 0,006 6 pF to 0,007 5 pF |
РПК 702-3 Direct method |
|||||
from 10 pF to 100 pF |
from 0,007 5 pF to 0,065 pF |
||||||
from 100pF tо 1000pF |
from 0,065 pF to 0,32 pF |
||||||
from 1000 pF to 10nF |
from 0,32 pF to 0,006 5 nF |
||||||
from 10 nF to 100 nF |
from 0,006 5 nF to 0,065 nF |
||||||
from 100 nF to 1 µF |
from 0,065 nF to 0,0058 µF |
||||||
At a frequency 300 Hz |
|
||||||
from 1 nF to 10 nF |
from 0,023 nF to 0,064 nF |
РПК 702-3 Direct method |
|||||
from 10 nF to 100 nF |
from 0,064 nF to 0,59 nF |
||||||
from 100 nF to 1 µF |
from 0,59 nF to 0,012 µF |
||||||
from 1 µF to 10 µF |
from 0,012 µF to 0,17 µF |
||||||
from 10 µF to 50 µF |
from 0,17 µF to 1,20 µF |
||||||
10.
|
Digital Frequency Meters (Frequency, Period, Time Interval Meters)
|
Frequency Hertz (Hz)
|
Frequency from 2 Hz to 1,5 GHz for a level no higher than 10 V реак tо реак |
|
|||
from 2 Hz to 1 000 Hz |
from 0,05 Hz to 0,01 Hz |
РПК 702-7 Direct method |
|||||
from 1 000 Hz to 1 MHz |
from 0,01 Hz to 0,000 0008 MHz |
||||||
from 1 MHz to 1 GHz |
from 0,000 0008 MHz to 0,000 0013 GHz |
||||||
from 1 GHz to 1,5 GHz |
from 0,000 0013 GHz to 0,000 0017 GHz |
||||||
Time interval second, (s)
|
Time interval from 1 µs to 1 000 ms for a level no higher than 10 V реак to реак |
|
|||||
from 1 µs to 1 ms |
from 0,06 µs to 0,000 36 ms |
РПК 702-7 Direct method |
|||||
from 1 ms to 1 000ms |
from 0,000 36 ms to 0,04 ms
|
||||||
Period, second, (s)
|
Period from 0,1 µs to 1 000 ms for a level no higher than 10 V реак to реак |
|
|||||
from 1 µs to 1 ms |
from 0,06 µs to 0,000 3 ms |
РПК 702-7 Direct method |
|||||
from 1 ms to 1 000ms |
from 0,000 3 ms to 0,004 ms |
||||||
11.
|
Pulse generators (time sources)
|
Time interval/ period second, (s)
|
Период
from 0,1 µs to 1 ms
from 1 ms to 1 000ms
Time interval
from 0,1 µs to 1 µs
from 1 µs to 1 ms
from 1 ms to 1000 ms
|
from 0,031 µs to 0,00036ms
from 0,00036 ms to 0,04 ms
from 0,031 µs to 0,031 µs
from 0,031 µs to 0,00036 ms
from 0,00036 ms to 0,04 ms |
РПК 702-20 Direct method |
||
Voltage Volt, (V) |
DC voltage |
|
|||||
from 0,01 V to 1 V |
from 0,000 077 V to 0,000 25 V |
РПК 702-20 Direct method |
|||||
from 1 V to 100 V |
from 0,000 25 V to 0,018 V |
||||||
12. |
Very stable standard frequency generators |
Frequency Hertz (Hz) |
0,1 МHz; 1 МHz; 5 МHz; 10 МHz |
5.10 -8
|
РПК 702-1 Direct method/ method of comparison by comparator |
||
13. |
Generators, synthesizers and other frequency sources |
Frequency Hertz (Hz)
|
from 2 Hz to 1,5 GHz for a level no higher than 10 V реак tо реак |
|
|||
from 2 Hz to 1 kHz |
from 0,05 Hz to 0,000 01 kHz |
РПК 702-18 Direct method |
|||||
from 1 kHz to 1 MHz |
from 0,000 01 kHz to 0,000 000 8 MHz |
||||||
from 1 MHz to 1 GHz |
from 0,000 000 8 MHz to 0,000 001 3 GHz |
||||||
from 1 GHz to 1,5 GHz |
from 0,000 001 3 GHz to 0,000 001 6 GHz |
||||||
14. |
Oscilloscopes with a frequency band up to 100 MHz |
Time interval/ period second, (s) |
from 0,1 µs to 1 ms |
from 0,012 µs to 0,12 ms |
РПК 702-6 Direct method |
||
from 1 ms to 1 s |
from 0,12 ms to 0,12 s |
||||||
Voltage Volt, (V) |
from 0,01 V to 5 V |
from 0,003 V to 0,13 V |
|||||
from 5 V to 25 V |
from 0,13 V to 0,7 V |
||||||
15. |
Gauge blocks
|
Length meter (m) |
from 0,5 mm to 100 mm |
from 0,000 15 mm to 0,000 26 mm |
РПК 702-5 method of comparison by comparator |
||
16. |
Micrometers |
Length meter (m) |
from 0,5 mm to 500 mm |
from 0,001 6 mm to 0,005 0 mm |
РПК 702-16 Direct method |
||
17. |
Caliper instruments (altimeters, depth gauges, calipers) |
Length meter (m) |
from 0,5 mm to 500 mm |
from 0,053 mm to 0,054 mm
|
РПК 702-15 Direct method |
||
18. |
Line standard metres -retractable pocket rules -measuring lines |
Length meter (m) |
from 0 mm to 10 000 mm
|
0,60 mm |
РПК 702-14 Direct method |
||
from 0 mm to 2000 mm |
0,60 mm |
||||||
19. |
Length indicators |
Length meter (m) |
from 0 mm to 25 mm |
0,003 mm |
РПК 702-17 Direct method |
||
20.
|
Vacuum gauges: mechanical with elastic-element and digital vacuum gauges |
Negative pressure kPa |
from minus 80 kPa to 0 kPa
|
from 0,10 kPa to 0,06 kPa
|
РПК 702-4 and РПК 702-2 Direct method |
||
Pressure gauges: mechanical with elastic-element and digital pressure gauges |
Positive pressure kPa and MPa |
from 0 kPa to 200 kPa |
from 0,06 kPa to 0,13 kPa |
||||
from 200 kPa to 2000 kPa |
from 0,13 kPa to 0,30 kPa |
||||||
from 2 MPa to 7 MPa |
from 0,000 3 МРа to 0,000 7 МРа |
||||||
from 7 MPa to 69 MPa |
from 0,0007 МРа to 0,008 МРа |
||||||
21. |
DC resistance meters: analog and digital (ohmmeters, ground testers, megaohmmeters, network testers, loop resistance testers) |
Electrical resistance Ohm (Ω) |
from 0,01 (Ω) to 100 (Ω) |
from 0,0023 (Ω) to 0,0033 (Ω) |
РПК 702-21 Direct method |
||
from 100 (Ω) to 10 (Ω) |
from 0,0033 (Ω) to 0,000 52 (Ω) |
||||||
from 10 (Ω) to 1 (Ω) |
from 0,000 52 (Ω) to 0,000 31 (Ω) |
||||||
from 1 (Ω) to 1 000 (Ω) |
from 0,000 31 (Ω) to 0,69 (Ω) |
||||||
22.
|
DC Resistance (single and multiple value measures) |
Electrical resistance Ohm (Ω)
|
from 0,1(Ω) to 100 (Ω) |
from 0,0012 (Ω) to 0,016 (Ω) |
РПК 702-11 Direct method |
||
from 100 (Ω) to 10 (Ω) |
from 0,016 (Ω) to 0,0023 (Ω) |
||||||
from 10 (Ω) to 1 (Ω) |
from 0,0023 (Ω) to 0,000 58 (Ω) |
||||||
from 1 (Ω) to 1 000 (Ω) |
from 0,000 58 (Ω) to 0,82 (Ω) |
||||||
23.
|
DC and AC clamp meters
|
Direct current Ampere (А) |
from 1 А to 100 А
from 100 А to 500 А
from 500 А to 1 000 А |
from 0,01 А to 0,20 А
from 0,20 А to 0,36 А
from 0,36 А to 0,68 А |
РПК 702-10 Direct method |
||
Alternating current Ampere (А)
|
|
|
|||||
At a frequency from 50 Hz to 100 Hz
from 1 А to 100 А
|
from 0,01 А до 0,21 А
|
РПК 702-10 Direct method |
|||||
from 100 А до 500 А |
from 0,21 А до 0,40 А |
||||||
from 500 А to 1 000 А |
from 0,40 А to 0,77 А |
||||||
24. |
Wattmeters |
Active power, 50 Hz Watt (W) |
cos φ = 1 |
|
|||
from 23 W to 460 W
|
from 0,10 W to 0,14 W
|
РПК 702-13 Direct method |
|||||
from 460 W до 2300 W
|
from 0,14 W to 0,47 W
|
||||||
cos φ = 0,5 positive
|
|
||||||
from 11,5 W to 230 W
|
from 0,54 W to 0,54 W
|
||||||
from 230 W to 1150 W
|
from 0,54 W to 0,62 W
|
||||||
cos φ = 0,5 negative
|
|
||||||
from 11,5 W to 230 W
|
from 0,54 W to 0,54 W
|
||||||
from 230 W to 1150 W |
from 0,54 W to 0,59 W |
References:
РПК 702-1 - Calibration methodology of very stable standard frequency sources
РПК 702-2 - Calibration methodology of digital pressure and vacuum gauges
РПК 702-3 - Calibration methodology of inductance and capacity meters
РПК 702-4 - Calibration methodology of elastic-element pressure and vacuum gauges
РПК 702-5 - Calibration methodology of gauge blocks by direct contact method of comparative measurement
РПК 702-6 - Calibration methodology of oscilloscopes
РПК 702-7 - Calibration methodology of digital frequency meters
РПК 702-8 - Calibration methodology of voltage calibrators
РПК 702-9 - Calibration methodology of current calibrators
РПК 702-10 - Calibration methodology of clamp meters
РПК 702-11 - Calibration methodology of single and multiple value resistances
РПК 702-12 - Calibration methodology of ammeters and voltmeters
РПК 702-13 - Calibration methodology of wattmeters
РПК 702-14 - Calibration methodology of line standard metres
РПК 702-15 - Calibration methodology of caliper instruments
РПК 702-16 - Calibration methodology of micrometric instruments
РПК 702-17 - Calibration methodology of length indicators
РПК 702-18 - Calibration methodology of frequency generators (frequency synthesizers)
РПК 702-20 - Calibration methodology of time sources
РПК 702-21 - Calibration methodology of electrical resistance meters
I HEREBY ORDER
the issuance of an Accreditation Certificate with ref. No. LC dated 8.12.2019 valid until 31.05.2021 with this Order as supplement and an integral part thereof.
Accreditation Certificate with the supplement thereof shall be obtained from the Director of DEFENCE INSTITUTE „PROFESSOR TSVETAN LAZAROV", the head of the Central
EA BAS Order No. А 594/18.12.2019
Laboratory for Measurement Equipment (CLME) at the Defense Institute “Prof. Tsvetan Lazarov" or another authorized person in the EA BAS building.
Upon receipt of the issued certificate and supplement thereof, the accredited person shall be obliged to return to EA BAS the originals of the accreditation certificate Ref. No. 7 LC dated 31.05.2018 and Annex - Order of EA BAS No. 182/31.05.2018.